Lampades Xenon Arcus Brevis Cermax®
Specificationes Operationales | ||
Descriptio | Nominalis | Spatium |
Potestas | CL Vatta | 100-150 Vatta |
Currens | 11 amperes (DC) | 10-14 amperes (DC) |
Tensio Operativa | 11.7 Voltia (DC) | 10-13.6 Voltia (DC) |
Tensio Ignitionis | 23 Kilovolts (ratio dependens) | |
Temperatura | 120℃ (Maximum) | |
Tempus Vitae | 1000 horae typicae |
Initialis Emissio ad Potentiam Nominalem | |
F = Egressus Filtratus UV | |
Descriptio | PE150AF |
Radius Emissus* | 16 Vatta |
Emissio UV* | 0.9 Watts |
Egressus IR* | Octo Vatta |
Egressus Visibilis* | 1350 Lumina |
Temperatura Coloris | 5900° Kelvin |
Instabilitates Maximae | 4% |
* Hi valores indicant emissionem totalem in omnibus directionibus. Longitudines undarum = UV<390 nm, IR>770 nm,
Visibilis: 390 nm-770 nm
* Finis vitae definitur ut 50% initialis producti
* Valores nominales ad 150 vattia post duas horas usus.
Descriptio | Exitus Visibilis | Productio Totalis* |
Apertura 3 mm | 700 Lumina | 5.5 Watts |
Apertura 6 mm | Lumina 900 | 8.0 Watts |
1. Lucerna non debet fenestra sursum versa intra 45° a verticali angulo operari.
2. Temperatura sigilli non debet excedere 150°.
3. Fontes potentiae regulati currenti/potentia et unitates receptaculorum lampadum Excelitas commendantur.
4. Lucerna intra limites commendatos currentiae et potentiae operanda est. Nimia potentia ad instabilitatem arcus, difficilem initium et praematuram senescentiam ducere potest.
5. Speculi calidi congregatio ad filtrationem infrarubram praesto est.
6. Lampades Cermax® Xenon multo tutiores sunt ad usum quam earum aequivalentes lampades arcus xenon quartz. Cautio tamen adhibenda est cum lampades operantur, quia sub alta pressione sunt, altam tensionem requirunt, temperaturas usque ad 200℃ attingunt, et radiatio infrarubra et ultraviolacea ustiones cutis et laesionem oculorum causare potest. Quaeso, Schedulam Periculi cum singulis lampadibus missis inclusam lege.